FUCHS Medical-Grade Grease: Lubrication Solutions Compatible with Both High-Temperature and Low-Temperature Sterilisation
Bearings, linear positioning assemblies, sliding components, connectors, seals and drug delivery devices in medical instruments are often challenged by friction and wear, material compatibility, assembly force and repeated sterilisation cycles. One major difficulty is that a single moving part may undergo different sterilisation processes sequentially, such as high-temperature steam sterilisation followed by low-temperature ethylene oxide sterilisation.
Lubricants must keep mechanisms operating reliably and continue meeting instrument design requirements after exposure to high-temperature moist heat or low-temperature chemical / radiation sterilisation. Leveraging the NYEMED product line, comprehensive testing capabilities and application engineering support, FUCHS helps customers develop lubrication solutions compatible with both high‑ and low-temperature sterilisation, spanning material screening through to process validation.

I. Why Sterilisation Processes Change Lubricant Selection
Unexpected friction or wear may occur at any stage of a medical device lifecycle: proof of concept, design verification, production ramp-up and field trials. Friction and wear degrade device performance. Beyond friction reduction, lubricants also serve sealing, motion control, damping and assembly aid functions.
When components undergo repeated sterilisation, grease is subjected to vastly different temperature and environmental conditions. High-temperature sterilisation involves humid heat and high pressure, while low-temperature sterilisation exposes lubricants to chemical media or radiation. Evaluating only initial torque or visual appearance at ambient temperature is insufficient. The sterilisation temperature range, sterilisation method, number of cycles, contacting materials, dispensing location and service-life requirements must all be included in validation assessments.
High-temperature sterilisation is typified by autoclaving, a demanding hot-wet service condition. Lubrication for low-temperature sterilisation compatibility most commonly targets ethylene oxide (EtO) processes; some applications also require gamma radiation assessment. The NYEMED range has undergone sterilisation stability testing for mainstream methods including autoclaving, ethylene oxide and gamma irradiation. (Test data helps narrow candidate selections; final confirmation of specific products, formulations, substrates and parameters must be performed on the actual device.) This supports early-stage selection for applications requiring compatibility with both high‑ and low-temperature sterilisation.

II. Key Validation Points for Lubricants Compatible with High‑ and Low‑Temperature Sterilisation
High-Temperature Steam Processes
Autoclaving combines heat, pressure and moisture. The primary risks for lubricants are thermal ageing, hydrolysis, oil separation and migration. Validation compares appearance, consistency, breakaway torque, operating noise and sealing performance before and after sterilisation, with focus on migration, volatilisation and interactions with substrate materials. Typical NYEMED data covers flash point, apparent viscosity, oil separation and evaporation, to support initial evaluation matrices.
Low-Temperature Ethylene Oxide and Gamma Radiation Processes
For low-temperature sterilisation, critical concerns include EtO permeation, residue removal and material adsorption. For gamma irradiation, testing focuses on whether grease viscosity, frictional properties and sealing performance degrade after exposure. Low-temperature conditions generally avoid the volatilisation risks of hot moist environments, so ageing mechanisms differ substantially from steam autoclaving. Multi-cycle testing matching real-world sterilisation protocols is recommended, with judgement criteria covering mechanism function, shelf life and packaging integrity.
*For dual-compatibility applications (both high and low temperature): Validation of a single sterilisation method alone is inadequate. Separate or alternating high-temperature steam cycles and EtO low-temperature cycles are recommended, to evaluate lubrication performance degradation after repeated exposure to mixed sterilisation modalities.
III. FUCHS: Product and Technical Support from Development through Serial Production
FUCHS has partnered with medical device and biopharmaceutical manufacturers for more than 30 years, providing support at every phase of customer product development. For devices subject to both steam autoclaving and EtO low-temperature sterilisation, customers are invited to share component materials, motion type, sterilisation temperature profile, sterilisation method and target service life. FUCHS application engineers will create custom validation plans and recommend suitable products and test protocols to implement lubrication solutions compatible with dual high‑ and low-temperature sterilisation.
If you have related requirements, please contact us via the official website enquiry channels for further discussion.